LABPEAKSpectrophotometers & Colorimeters
Atomic Absorption Spectrophotometer
Specifications
- Model
- SPL-PAA4530
- Wavelength Range
- 190-900nm
- Size (W×D×H)(mm)
- 730×625×700
- Accuracy of the Wavelength
- ±0.15nm
- Receptivity of the Wavelength
- ≤0.04nm
Full specification sheet
- Model: SPL-PAA4530
- Wavelength Range: 190-900nm
- Spectral Bandwidth: 0.1/0.2/0.4/1.0/2.0 nm
- Accuracy of the Wavelength: ±0.15nm
- Receptivity of the Wavelength: ≤0.04nm
- Base Line Stability: ≤0.002A/30min (Cu)
- Characteristic Viscosity: 0.02μg/ml/1% (Cu)
- Checking Limit: 0.004μg/ml (Cu)
- Preciseness: 0.005
- Grating: 1800 lines/mm
- Inflamer: All-metal titanium burner
- Atomizer: Effective glass atomizer
- Lamp Stand: 8
- D2 Background Correction Ability: When the background is 1A, the background ability should be deducted not less than 50 times, self absorption background deduction method
- Power Source: 220V±22V AC
- Shipping Size(W×D×H)(mm): 860×705×755(Main instrument)
- Shipping Size(W×D×H)(mm): 1170×645×900(Accessories)
- G.W. (kg): 105 (Main instrument); 100 (Accessories)
- G.W. (kg): Flame System
- Acetylene Air Burner: 100mm
- Lgnition Dynamic Baseline Drift: ≤0.006A/30min
- (Cu) Characteristic Viscosity: ≤0.025μg/ml/1%
- Related Standard Deviation of the Accuracy: ≤0.5% (Cu, absorbance>0.8A) (detection limit Cu≤0.008μg/ml)
- Safety System: Can automatically cut off the gas when the pressure is not enough the power is off flame out and unconformity of the burner
- Safety System: Furnace
- The Highest Temperature: 3000℃
- The Largest Temperature Rising Speed: ≥2000℃/s
- Characteristic Quantity: Cd≤0.5×10-12g Cu≤0.5×10-11g
- Accuracy: Cu≤3% Cd≤3%
- Size (mm) and Weight: 730×625×700; 79.3kg
- Safety System: Over current protection
- Safety System: Low air pressure alarm/protection
- Safety System: Low cooling water flow alarm/protection
- Power Source and Power: 220V±22V AC 7000W